Supplement for Infinity Delta Series Software
28
Supplement – Infinity
®
Delta Series Software VF10.1
NBP Setup
To connect the hose
Push the hose end fitted with the plastic collar firmly into the connector on the left side of the monitor
(see figure below).
Venous Stasis
Available in software levels equal to or greater than VF9.1, and supports venous stasis in adult mode only
(with adult-based inflation pressures).
NOTE
The accuracy of the oscillometric blood pressure signal can decrease (up to loss of measurement)
under the following conditions:
– Weak pulses
– Irregular pulses
– Patient movement artifacts
– Tremor artifacts
– Respiratory artifacts
–
Pulses generated from a ventricular assist
device
These conditions may prevent the monitor from performing a measurement within the maximum
allowable measurement time.
NBP hose connector
Kappa
Delta/Delta XL
NBP hose connector