Supplement – Infinity
®
Delta Series Software VF10.1
87
Supplement for Infinity Delta Series Software
Immunity test
IEC 60601 Test Level:
Compliance
Level:
Recommended Separation
Distance:
Conducted RF
rf coupled into lines
(IEC 61000-4-6)
150 kHz to 80 MHz
RMS
See note 3.
Radiated rf
(IEC 61000-4-3)
80 MHz to 800 MHz
800 MHz to 2.5 GHz
3 V/m
where
P
is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and
d
is
the recommended separation
distance in metres (m).
1
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the equipment is used exceeds the applicable RF compliance level above, the
equipment should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the equipment.
2
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
3
Applicable for MultiGas Monitor (SCIO) over frequency range 150 KHz to 2.5 GHz and less than
1 V/m above 2.5 GHz.