Supplement for Infinity Delta Series Software
40
Supplement – Infinity
®
Delta Series Software VF10.1
Connection
Hardware - Module Connection
Connection ports are marked on the front of the Dräger mainstream CO
2
module. Use the diagram below
as a guide when connecting the module, cables, and accessories to the monitor.
CAUTION
To prevent leakage, make sure the cuvette is firmly connected to the breathing circuit.
CAUTION
Check the CO
2
mainstream sensor and measureing cuvette for damage before use. A damaged CO
2
sensor may impair electrical isolation or may introduce debris into the breathing circuit.
CAUTION
Only use the mainstream CO
2
sensor with the approved CO
2
measuring cuvettes for adult or pediatric
applications. See the “Approved options and accessories” section. The measuring function may
otherwise be inaccurate.
NOTE
In this chapter, all Infinity Delta Series patient monitors are referred to as “patient monitor.”
NOTE
When
etCO2
,
inCO2
or
RRc
is followed by * (for example,
etCO2*
), the * indicates the origin of this
parameter is from an external device such as an A500, A350, or A300.