Supplement for Infinity Delta Series Software
68
Supplement – Infinity
®
Delta Series Software VF10.1
When a MEDIBUS.X device is in standby, the Delta series monitor does not process parameters, settings,
waveforms, or modes (i.e., this data is not displayed on the Delta series monitor and is not sent to the
Infinity network).
Parameter values acquired from peripheral devices are trended in graphical and tabular format. When you
modify the parameter display, a message identifies the trend data that will be lost if the trend storage
capacity is reached.
In the list of parameters available from specific peripheral devices later in this chapter, an asterisk (*)
identifies parameters that duplicate the monitor's parameters.
Consult device documentation for details on setup and operation.
Compatible devices
The following table lists the devices supported by the Delta series monitors.
NOTE
Delta series monitors support Medibus.X version 1.07.
MIB II Duo Protocol Converter
(PN: 72 56 949)
MIB II Protocol Converter
(PN: 72 56 931)
Input: RS232, 25 Pin D
Input: RS232, 25 Pin D
Output: RJ45 MIB
Output MIB II: RJ45 MIB
Output MIB I: SDL
Device
MIB Interface Protocol
MEDIBUS.X
Interface Protocol
Dräger Anesthesia devices
Infinity Perseus A500
No
Yes
Atlan A350/A300
No
Yes
Dräger Apollo
Yes
No
Dräger Fabius Family
Yes
No
Dräger Primus
Yes
No