Supplement for Infinity Delta Series Software
To dry the sidestream pump subsystem
1
Reattach the etCO2 module or pod to the monitor. The sidestream sampling pump starts running, and
there is suction at the input port on the face of the module or pod.
2
With the input sidestream port still open and the drain tubing still connected, let the pump run for
several minutes to remove any water still trapped in the system.
3
Block the sidestream input port with your finger for several seconds and then unblock it. Repeat at
least ten times.
4
Move your finger to the sidestream output port and block the port with your finger for several seconds
and then unblock it. Repeat at least ten times.
5
Remove the drain tubing, and allow the sidestream pump to continue running for at least 30 minutes.
Technical Data
Electromagnetic Compatibility (EMC)
Electromagnetic Immunity
NOTE
If the sidestream pump fails to start, make sure the Capnostat sensor is disconnected. The pump is
designed to shut down while a connected sensor is warming up.
NOTE
Sidestream accessories are for single use only. Do not reuse.
This equipment is intended for use in the electromagnetic environment specified below. The
user of this equipment should assure that is used in such an environment.
Immunity against IEC 60601-1-2 Test Level: Compliance level
(of this device:
Electromagnetic Environment:
Electrostatic
discharge, ESD
(IEC 61000-4-2)
Contact discharge: ±6 kV
air discharge: ±8 kV
±6 kV
±8 kV
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30% so that electrostatic
charges are at suitable levels.
Electrical fast
transients / bursts
(IEC 61000-4-4)
Power supply lines: ±2 kV
longer input / output lines:
±1 kV
±2 kV
±1 kV
Mains power quality should be
that of a typical commercial or
hospital environment.
Surges on AC
mains lines
(IEC 61000-4-5)
Common mode: ±2 kV
Differential mode: ±1 kV
±2 kV
±1 kV
Mains power quality should be
that of a typical commercial or
hospital environment.