Supplement – Infinity
®
Delta Series Software VF10.1
11
Supplement for Infinity Delta Series Software
Temperature Display
All temperature readings appear on the main screen according to their position in Parameter Priority (see
page 2-6). The following display conventions govern temperature labels and values.
The monitor displays temperature values in one parameter box for each pod connector (MultiMed, He-
mo2/Hemo4). The variables “a” and “b” denote the first or second probe connector from the MultiMed with
Y-cable, NeoMed, Hemo2, Hemo4 pod or the MPod – Quad Hemo. When acquired via the MultiMed or
NeoMed pod, temperature signals are displayed as Ta <value> or Tb <value>. (If you are using a MultiMed
without a Y-cable, only Ta displays a value.)
When monitoring temperature using the Hemo2, Hemo4 pod, or the MPod – Quad Hemo, temperature
values are further identified according to where they are connected to the monitor. The temperature value
corresponding to the device hooked up to the Aux/Hemo2 PodCom connector is labeled T2a or T2b, while
the temperature value corresponding to the device hooked up to the Aux/Hemo3 connector is labeled T3a
or T3b.
NOTE
The transient response time for the monitor is less than 2.5 seconds.