Supplement – Infinity
®
Delta Series Software VF10.1
95
Supplement for Infinity Delta Series Software
Invasive Blood Pressure
Mode of operation
Continuous
Protection against electric
shock
Sensor: Type BF
Defibrillator protection
Yes
Environmental requirements
Temperature range
Operating: 10 to 40
°
C (50 to 104
°
F)
Storage: –40 to 75
°
C (–40 to 167
°
F)
Relative humidity
Operating: 20 to 90%, non-condensing
Storage: 10 to 95% (with packaging)
Atmospheric pressure
Operating: 525 to 795 mmHg (70 to 106 kPa)
Storage: 86 to 825 mmHg (11.5 to 110 kPa)
Physical specifications
Frequency ranges
DC to 8, 16, or 32 Hz ± 10% (user selectable)
Accuracy
±1 mmHg or ±1% exclusive of transducer (whichever is greater)