Supplement for Infinity Delta Series Software
96
Supplement – Infinity
®
Delta Series Software VF10.1
Monitoring Specifications
ECG
Respiration
Pulse Contour Cardiac Output (PiCCO)
WARNING
The following parameters and functions are not monitored in neonatal mode: Arrhythmia,
cardiac output, ST, pacer detection,% paced, and all MultiGas parameters.
Description
Value
V/s pacer detected
3.18 V/s
Description
Value
Detection threshold:
0.15
Ω
to 4.0
Ω
in manual mode (user adjustment)
Adult and Pediatric: 0.20
Ω
- 1.5
Ω
in auto mode
(automatic adjustment)
Neonatal:
0.20
Ω
- 1.5
Ω
in auto mode
(automatic adjustment)
Measurement range:
Range: 0 - 154 breaths per min
Additional technical data
Measurement method
PCCI
Continuous Pulse Contour Analysis
PCCO
Continuous Pulse Contour Analysis
p-SVI
Continuous Pulse Contour Analysis
SVV
Continuous Pulse Contour Analysis
p-SVR
Continuous Pulse Contour Analysis
p-SVRI
Continuous Pulse Contour Analysis
dPmax
Continuous Pulse Contour Analysis
GEF
Transpulmonary Thermodilution
p-CO
Transpulmonary Thermodilution
GEDV
Transpulmonary Thermodilution
GEDVI
Transpulmonary Thermodilution
EVLW
Transpulmonary Thermodilution
EVLWI
Transpulmonary Thermodilution
PVPI
Transpulmonary Thermodilution
Display resolution
PCCI
0.01 L/min/m
2
PCCO
0.01 L/min
p-SVI
1 ml/m
2
SVV
1%
p-SVR
10 dyn.sec. cm
-5
p-SVRI
10 dyn.sec. cm
-5
/m
2
dPmax
1 mmHg/s
GEF
1%
p-CO
0.01 L/min
GEDV
1 ml
GEDVI
1 ml/m
2
EVLW
1 ml
EVLWI
1 ml/kg
PVPI
0.1
p-BT
0.1 ºC (1 ºF)