Supplement – Infinity
®
Delta Series Software VF10.1
107
Supplement for Infinity Delta Series Software
Output signals
Graphical signal of the output voltage and current stimulation pulse:
A
– Voltage (V
off
) or Current (I
off
) when pulse is not delivered (OFF)
B
– Voltage (V
on
) or Current (I
on
) when pulse is delivered (ON)
C
– Pulse width (varies, dependent on setting)
1
Stimulation current accuracy is ±5% within this range. Stimulation current is disabled at
impedance values greater than 8900
Ω
.
2
At a stimulation current of 30 mA, voltage and current values are approximately half (50%) of
those listed for 60 mA.
NOTE
: There is no output in an open-circuited load condition.
A
B
C
Output voltage and current at stimulation current of 60mA
(2)
Load impedence 1.2
Ω
500
Ω
1000
Ω
6000
Ω
V
on
0.07 V
29.2 V
60.8 V
342.0 V
V
off
0 V
0 V
0 V
0 V
I
on
57.5 mA
59.8 mA
59.3 mA
57.0 mA
I
off
0 mA
0 mA
0 mA
0 mA