Auxiliary Air Valve of the Patient System
The auxiliary air valve allows the patient to spontaneously breathe ambient air should
the medical gas supply and/or Fabius fail. The opening pressure of the auxiliary air
valve is 0 to –5 hPa (mbar).
Fig. 32:
Sectional view of the auxiliary air valve
threaded ring
valve seat
valve disc
valve cross with spring