• Remove fixing screws
1
of S-ORC and mount S-ORC to new flowmeter block.
Fig. 92:
S-ORC, removing the fixing screws
• Connect marked hoses to S-ORC by pushing them firmly into hose connections.
• Connect marked hoses of Fabius to new flowmeter block by pushing them firmly
into hose connections.
• Check that hoses are properly connected by pulling them without using too much
force.
• Mount shield to new flowmeter block.
• Mount new flowmeter block to Fabius.
• Check hose connections of flowmeter block and of S-ORC for leaks.
• Check function of flowmeter block using the Test Certificate.