Spontaneous Breathing: Expiration
During expiration, the inspiratory valve
1
remains closed thus preventing the
expiratory gas from flowing back into the inspiratory branch.
The APL valve
2
is open, irrespective of its pressure setting.
The expiratory gas flows from the lung
3
through the expiratory hose
4
, the flow
sensor
5
, the PEEP control valve
6
, the expiratory valve
7
, and through the absorber
8
into the breathing bag
9
. At the same time, new fresh gas flows into the breathing bag.
When the breathing bag is full, any excess gas mixture flows through the non-return
valve
10
into the anesthetic gas scavenging system.
Fig. 16:
Spontaneous breathing (expiration)
The CO
2
is scrubbed from the expiratory gas by the soda lime contained in the
absorber
8
. The fresh gas replaces the anesthetic and oxygen taken up by the
patient.
2
10
9
3
8
fresh gas
1
7
5
6
SPONT/
IPPV
4
Summary of Contents for Fabius 4116029
Page 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...