Fabius
®
Repair Instructions
3-132
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_
repa
ir
instr
uc_c.
fm
• Remove bag-type rolling seal
1
by pulling it out of ventilator in direction of
A
.
Fig.125:
View of the dismounted ventilator, removing the bag-type rolling seal
• Disconnect hose from connection port
2
.
• Remove fixing screws
3
from cover.
Fig.126:
Side view of the ventilator, removing the fixing screws
1
A
3
2