Fabius
®
Repair Instructions
3-51
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_r
epai
rinstruc_c.fm
• Remove hose
1
from pressure sensor (Paw).
• Remove fixing screws and washers
2
of pressure sensor (Paw).
• Carefully remove pressure sensor and mount it to new Control PCB.
• Remove EPROM
3
from old Control PCB and mount it to new Control PCB
(
make sure to mount EPROM correctly (direction) into EPROM mount)
.
• Cut cable tie
4
which holds rechargeable battery in place.
• Remove rechargeable battery and mount it to new Control PCB using a new
cable tie (
make sure to mount rechargeable battery correctly (direction)
.
• To avoid wrong connections, mark hoses
5
of flow sensor.
• Disconnect hoses
5
from flow sensor.
Fig. 33:
Removing pressure sensor (Paw), differential pressure sensor, rechargeable
battery and EPROM
D 3
B 3
X13
X23
X16
X4
X6
X1
X8
R40
D7
G 1
D5
L1
L2
X3
+
D
2
1
3
5
5
4
Summary of Contents for Fabius 4116029
Page 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...