Intermittent Positive Pressure Ventilation: Expiration
During expiration, the inspiratory valve
1
remains closed thus preventing rebreathing
into the inspiratory branch.
The expiratory gas flows through the expiratory valve
2
, the flow sensor
3
, back into
the breathing bag
4
and, at the same time, mixes with the fresh-gas from the fresh
gas port
5
.
The ventilator piston moves back drawing the gas mixture needed for the next
inspiration into the piston space.
The excess fresh gas flows through the APL valve
6
, and through the non-return valve
7 in
to the anesthetic gas scavenging system
8
.
Fig. 25:
Intermittent positive pressure ventilation (expiration)
Intermittent Positive Pressure Ventilation: Expiration with PEEP
A PEEP value is adjusted on the control box. The corresponding PEEP control
pressure is applied to the diaphragm of the PEEP valve. The diaphragm plate of the
PEEP valve pushes the mica disc of the expiratory valve which closes the crater. If the
pressure of the expiratory gas exceeds the set PEEP value, the expiratory valve
2
opens.
1
5
3
2
7
8
4
6
ventilator
PEEP valve
Summary of Contents for Fabius 4116029
Page 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...