Fabius
®
Repair Instructions
3-145
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_r
epai
rinstruc_c.fm
• Disconnect cable connectors
1
from Pneumatics Control PCB.
• Remove fixing screws
2
.
Fig.143:
Removing the fixing screws from the Pneumatics Control PCB
• Replace the Pneumatics Control PCB with a new one.
• Secure new Pneumatics Control PCB to pneumatics using fixing screws.
• Connect cable connectors
1
to Pneumatics Control PCB.
• Mount pneumatics to Fabius using reverse method of that used for removal.
• Measure equivalent device leakage current according to Test Certificate.
• Measure protective conductor resistance according to Test Certificate.
• Check function of Pneumatics Control PCB according to Test Certificate.