23.2.2
Safety Valve Components
Fig.160:
Sectional view of the safety valve
23.2.3
Dismounting/Replacing the Safety Valve
• Unscrew safety valve using a pair of plastic-jaw pliers.
• Remove residual Loctite from safety valve mount on patient system.
• Dispose of the old safety valve.
• Apply a small amount of Loctite #222 onto thread of the new safety valve and
screw safety valve into patient system using plastic-jaw pliers.
Screw
Valve disc
Disc
Spring