Fabius
®
Repair Instructions
3-120
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_
repa
ir
instr
uc_c.
fm
• Disconnect hoses from hose connections
1
by pushing in the retaining rings and,
at the same time, disconnecting the respective hose.
• Remove S-ORC fixing screws.
• Replace the S-ORC with a new one.
• Mount new S-ORC to flowmeter block.
Fig.107:
Side view of the flowmeter block (left)
• Connect previously marked hoses to S-ORC.
• Connect previously marked hoses to flowmeter block.
• Mount flowmeter block using the reverse method of that used for dismounting.
• Check S-ORC’s function according to relevant Test Certificate.
;
S-ORC
140 lg
140 lg
200 lg
200 lg
N
2
0
0
2
A E
E
A
Flowmeter block
S-ORC
1
1
1