Pressure Regulators
Replace pressure regulators after six years of service or install “Pressure Regulator”
spare parts set. See replacement interval matrix on Page 4.5 of the Test Certificate.
15.4.1
Testing the Pressure Regulators
• Check pressure regulators according to Test Certificate.
15.4.2
Mounting the "Pressure Regulator" Spare Parts Set
• Switch control box off using ON/OFF switch (ON/OFF switch is located on rear
panel of control box).
• Disconnect power plug from AC outlet.
• Disconnect gas supply hoses from pipeline system.
• Remove gas box (see "
• Mark hoses of respective pressure regulator for later reassembly, for example,
1
for outlet,
2
for inlet, in order to avoid interchanging of inlet and outlet of pressure
regulator
(note the arrow marking on pressure regulator)
.
Fig. 69:
Marking pressure regulator hoses
O2 pressure regulator
N2O pressure regulator
AIR pressure
1
2
Summary of Contents for Fabius 4116029
Page 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...