rinstruc_c.fm
• Connect connector of piston pump to connection
1
of Pneumatics Control PCB.
• Connect hoses
2
and
3
(see previously applied markings) to piston pump
4
(see
following Figure).
• Fasten electrical cables together using a new cable tie.
• Cut off any protruding length of cable tie.
Fig.148:
Top view of the pneumatics, connecting the piston pump
• Mount pneumatics to Fabius using reverse method of that used for removal.
• Measure equivalent device leakage current according to Test Certificate.
• Measure the protective conductor resistance according to Test Certificate.
• Check function of piston pump according to Test Certificate.