Function Principle of the Pneumatic Control of the Ventilator
The piston pump generates a vacuum required for the bag-type rolling seal and the
rolling seal of the ventilator as well as the control pressure for the PEEP valve. The
spring-loaded non-return valve limits the vacuum to
−
200 hPa (mbar). A pressure
sensor measures the current pressure and converts the pressure value into a
corresponding electrical signal. This signal is transmitted to the microprocessor. The
reservoir V1 "smooths" the flow to the electronic PEEP valve. The combination of filter
and reservoir V2 dampen the noise.
Fig. 28:
Function diagram of the pneumatic control of the ventilator
ventilator
E
P
piston
pump
–200 hPa (mbar)
reservoir V1
electronic
PEEP valve
to the PEEP
valve/ expiratory
valve
filter
reservoir V2
pneumatics
Summary of Contents for Fabius 4116029
Page 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...