4-24
21.3
Calibrate pressure sensor
In Service mode set pressure sensor to
zero. (Keycode 112)
Detach pressure gauge tube to do so.
Calibration successful.
(Otherwise replace pressure sensor 18 31
887)
Switch unit off and back on again.
Set to Man/Spont.
Connect bellows or disposable syringe
and pressure gauge (-30 - +120 mbar) to
pressure gauge tube.
Build up pressure 40 mbar + 0,5 mbar.
Display “P
mean
” rises to 40 + 2
Remove test set-up.
Connect P
AW
an hose to control box.
21.4
Set O2 sensor to zero
Call up zeroing page by pressing
p
Calibrate O2 sensor with 100 % by vol. O2
Attach insp. O2 sensor to test adapter,
item no. 7900138.
Connect fresh-gas outlet to test adapter.
Flush with approx. 100 % O2 (1 L/min.) for
approx. 1 minute.
Then trigger calibration.
Calibration is completed after max. 4
minutes.
21.5
O2 sensor test:
Switch to MAN/SPONT. Expose sensor to
ambient air for approx. 2 minutes
18 - 24 % by vol. O2 is displayed on
completion of this period.
If not, replace sensor capsule.
Plug in O2 sensor
Remove test set-up.
O
V
O
O
Summary of Contents for Fabius 4116029
Page 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...