Pressure Sensor (Paw) on Control PCB
In case of failure, replace the complete pressure sensor (Paw) on the Control PCB.
After replacement, calibrate the pressure sensor (Paw) in the DrägerService Mode.
14.14.1 Testing the Pressure Sensor (Paw)
• Test pressure sensor (Paw) in the debug mode: “P” and “Plat”.
14.14.2 Dismounting/Replacing the Pressure Sensor (Paw)
• Switch control box off using ON/OFF switch (ON/OFF switch is located on rear
panel of control box).
• Remove control box from Gas Box (see "Removing the Control Box from the Gas
Box").
• Remove lateral screws
1
from control box
2
and lift top cover
3
in direction of
A
and away from control box
2
.
• Place top cover
3
of control box
2
aside.
Fig. 35:
Front view of control box, removing top cover
Electrostatic discharge can damage electrostatic sensitive
devices. Use a static-dissipative mat and a wrist strap when
handling electrostatic sensitive devices.
1..2..
Menu
PEEP
IPPV
man./
spont.
Pmax
TI:TE
TIP:TI
f
IPPV
V
T
IPPV
Paw [mbar]
FiO2
MV
Paw
60
40
20
0
24
7.9
28
0
80
21
12.0
3.0
40
8
1
1
A
2
3
Summary of Contents for Fabius 4116029
Page 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...