General Information About the Uninterruptible Power
Supply (Option)
The uninterruptible power supply (rechargeable batteries) must be replaced with a
new one at 3-year intervals. In case of malfunction, the only replaceable single part is
the fuse of the rechargeable batteries.
21.1
Dismounting/Replacing the Uninterruptible Power Supply
• Switch control box off using ON/OFF switch (ON/OFF switch is located on rear
panel of control box).
• Unplug power plug from AC outlet.
• Remove fixing screws
1
from cover plate
2
(rear panel of Fabius) and place cover
plate
2
aside.
Fig.137:
Rear view of Fabius, removing the uninterruptible power supply
1
1
2