4-16
11.9
Pressure gauge
Comparison measurement
Tolerance
±
4% of full scale value
11.10
Shutoff valve
11.11
Vacuum control valve (vent valve)
11.12
Pushbutton rapid vent valve
(vacuum operation only)
12.
AGS system (optional)
If applicable, replace particulate filter
M 33294
When using one of the two surplus gas
connections:
Plug fitted in socket on side of AGS
mounting system.
When using both surplus gas
connections:
Plug removed from socket on side of AGS
mounting system.
AGS functional check
Press plug into outlet point of anaesthetic
extractor until it engages.
Indicator turns green.
Float in flowmeter of AGS system moves
between upper and lower mark.
Detach extractor from wall socket.
13.
Vapor plug-in system housing section
Renew O-ring U 4314 (2 x each Vapor)
once a year
Date of next replacement: _________
14.
Safety valve, flowmeter unit
Perform test once a year.
Connect mixed-gas socket to precision
pressure gauge +1.0 bar, 7901513.
Uncouple Vapor.
N2O control valve closed.
Set
*
= 9 L/min. at O2 flow control valve.
Test value:
Safety valve opens at p = < 1.0 bar
C O
O L
O L
O L
C
*
C
C
O
O
*
C
O
Summary of Contents for Fabius 4116029
Page 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...