4-20
17.2
Testing of inhalation and exhalation
valves
Remove adapter. Seal absorber with
protective cap (M 14405). Remove
manual breathing bag. Connect up
pressure gauge
-30/+120 mbar, rotameter as shown.
Outlet valve function test:
seal
SEAL
FRESH-GAS
DECOUPLING
VALVE
SEAL
SV00421
EXPIRATION VALVE