19.2.5
Checking/Adjusting the O
2
Control Pressure
• Connect a pressure gauge (measuring range 100 mbar) to O
2
side inlet port
1
using a T-piece.
• Adjust O
2
flow control valve to 0.20 L/min.
The control pressure should be 25 mbar (±1 mbar).
• If the control pressure is not 25 mbar (±1 mbar) use adjusting screw
2
to adjust
the control pressure on the O
2
side.
Fig.102:
S-ORC, adjusting the O
2
control pressure