Repair Instructions
3-125
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_r
epai
rinstruc_c.fm
• Disconnect ventilator connector
1
from uninterruptible power supply.
• Push ventilator connector back into ventilator compartment.
Fig.113:
Rear view of Fabius, disconnecting the ventilator connector
• Hold patient system in place (to prevent ventilator from falling down) and remove
fixing screw and washer
2
.
Fig.114:
Top view of Fabius, removing fixing screw and washer
• Place ventilator on a stable surface.
20.4
Dismounting/Replacing the Motor
In case of failure, replace the complete ventilator with a new one.
1