Fabius
®
Repair Instructions
3-52
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_
repa
ir
instr
uc_c.
fm
• Remove fixing screws
1
.
Fig. 34:
Control PCB, removing fixing screws
• Replace Control PCB with a new one.
• Secure new Control PCB to control box using fixing screws
1
.
• Connect hoses to flow sensor of new Control PCB
(make sure to connect
hoses as marked previously)
.
• Connect connectors.
• Secure O
2
sensor port using nut (
do not forget washer on inside of O
2
sensor
port)
.
• Mount top cover to control box.
• Mount control box to Fabius trolley.
• Connect cables and hoses to control box (see markings).
• Carry out a system initialization (system init, keycode 112).
• Enter previously recorded menu settings.
D 3
B 3
X13
X23
X16
X4
X6
X1
X8
R40
D7
G 1
D5
L1
L2
X3
+
D
1
1
1
1
1
1
In newer units, the
9V rechargeable
battery has been
replaced with a
Supercap capacitor.
Summary of Contents for Fabius 4116029
Page 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...