Fabius
®
Repair Instructions
3-61
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_r
epai
rinstruc_c.fm
14.16.4 Rechargeable Battery Connector X11
The following voltages should be measured according to the current operating mode
(battery-powered, AC-powered, or ON/OFF switch off): on-load voltage, charging
voltage, off-load voltage.
14.16.5 Display Backlighting Connector X4
Approximately 1000 VAC are applied at off-load, approximately 280 V AC are present
during operation.
14.16.6 Connector X8 (connection to Power Supply PCB)
During operation, +5 V are measured across red (+5V) and black wire (GND), and
approx. +27 V across black (GND) and blue wire (motor).
i
Voltage dips slightly during measurement.