Fabius
®
Repair Instructions
3-155
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_r
epai
rinstruc_c.fm
• Remove fixing screws
1
from pneumatics
2
.
• Loosen screw
3
.
• Lift pneumatics
2
up and pull pneumatics slightly out (you can latch pneumatics
onto the plates for convenience.)
Fig.154:
Rear view of Fabius, removing the pneumatics
Electrostatic discharge can damage electrostatic sensitive
devices. Use a static-dissipative mat and a wrist strap when
handling electrostatic sensitive devices.
1
1
2
3