• Calibrate pressure sensor (Paw) in the DrägerService Mode.
• Calibrate flow sensor in the debug mode.
• Re-calibrate PEEP valve.
• Calibrate oxygen concentration in the debug mode.
• Check Control PCB function according to Test Certificate.