Fabius
®
Repair Instructions
3-130
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_
repa
ir
instr
uc_c.
fm
• Turn motor upside down.
• Disconnect cable connector from motor.
• Remove screws
1
from motor.
Fig.122:
Bottom view of the motor, removing the fixing screws
• Replace the motor with a new one.
• Mount motor using the reverse method of that used for dismounting
making sure
the cables are installed correctly
.
• Connect cable connector
2
to motor.
Fig.123:
Top view of the ventilator, connecting the cable connector
• Mount ventilator to Fabius using the reverse method of that used for disassembly.