COMPASS
for molbox™ User’s Manual
Page 73
© 1998-2003
DH Instruments, Inc
.
FEATURE
DESCRIPTION
Set Range
Set Units
(conditional drop
down list/
text entry field)
Select the DUT set point unit from the drop down list. If the unit is not in the list, type the
unit into the list box. If the set range unit is the flow range unit, the min/max set point
fields are automatically set to the min/max flow range values. If the DUT’s
<Set Point
Type>
in the DUT Definition
[Set]
tab is
<Voltage/Current>
, the set point unit must be V,
mV, A or mA.
There is no set range if the DUT is NOT a controller.
Set Range
Min Set
(conditional
numeric entry
field)
Enter the lowest set point (set point corresponding to the
<Min Flow>
) of the DUT in the
set point unit. If the set point unit is the flow range unit, the
<Min Flow>
value is copied
here and cannot be changed.
There is no set range if the DUT is NOT a controller.
Set Range
Max Set
(conditional
numeric entry
field)
Enter the highest set point (set point corresponding to the
<Max Flow>
) of the DUT in the
set point unit specified above. If the set point unit is the flow range unit, the
<Max Flow>
value is copied here and cannot be changed.
There is no set range if the DUT is NOT a controller.
8.5.5 [TOLERANCE]
TAB
The DUT Definition
[Tolerance]
tab is used to provide a specification for the DUT. The DUT
specification is used in calculating DUT errors and in determining the status of a test point
when tests are run (see Section 7.2), the state of the
<Tolerance Indicator>
on the
<DUT/Ref
Comparison>
run screen (see Section 6.7.1) and is applied to test data plots to determine the
tolerance bars on plots (see Section 6.7.6). There are several tolerance formatting options.
Choose the selection appropriate for the DUT.
Table 19 provides details on each of the
[Tolerance]
tab fields.
%Rdg only tolerances should be avoided if test points at or near 0 will be taken.
The tolerance becomes infinitely small as the reference flow approaches zero. Instead, use a
combination tolerance such as <%DUTFS OR %Reading (Greater of)>.
Figure 42
. DUT Definition Editor,
[Tolerance]
Tab
Summary of Contents for COMPASS FOR MOLBOX
Page 1: ...1998 2003 DH Instruments Inc COMPASS for molbox Flow Calibration Software User s Manual...
Page 16: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 2 N NO OT TE ES S...
Page 20: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 6 N NO OT TE ES S...
Page 64: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 50 N NO OT TE ES S...
Page 94: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 80 N NO OT TE ES S...
Page 112: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 98 N NO OT TE ES S...
Page 150: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 136 N NO OT TE ES S...
Page 166: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 152 N NO OT TE ES S...
Page 172: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 158 N NO OT TE ES S...
Page 192: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 178 N NO OT TE ES S...
Page 194: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 180 N NO OT TE ES S...
Page 200: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 186 N NO OT TE ES S...
Page 216: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 202 N NO OT TE ES S...
Page 220: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 206 N NO OT TE ES S...
Page 224: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 210 N NO OT TE ES S...
Page 226: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 212 N NO OT TE ES S...