COMPASS
for molbox™ User’s Manual
Page 51
©
1998-2003
DH Instruments, Inc
.
7
7
.
.
[
[
R
R
U
U
N
N
]
]
M
M
O
O
D
D
E
E
S
S
7.1 OVERVIEW
COMPASS
has four basic run modes:
<Idle>
,
[Run Test]
,
[Run molbox]
and
[Run DUT]
. Other than
<Idle>
, which indicates that
COMPASS
is in a state in which configuration changes can be made, each of
these run modes is found in the
[Run]
menu on the main program. Before activating any of the run modes,
make sure that
COMPASS
has been configured to read the instruments required by the run mode
(see Section 10.1). With the exception of the tests in which all entries are manual, every run mode involves
continuous instrument polling to update data on the run screens. Adjust these screens as need to customize
the
COMPASS
display (see Section 6.7). Sections 7.2 through 7.4 detail the
COMPASS
run modes.
7.2 [RUN], [RUN TEST]
This menu selection is used to select a DUT Definition execute a test defined by a Test Definition
(see Sections 8, 9). Test execution is divided into three sequential parts:
Test Initialization (see Section 7.2.1):
Sets up the molbox interface, selects the DUT and Test
Definition to run and enters user ID. Many of the sequential steps of test initialization can be turned
ON or OFF using the
[Tools]
,
[Options]
,
[Initialize Test]
tab (see Section 11.5).
Run Test (see Section 7.2.2):
Runs purge (if included) and leak test (if included) followed by the
test flow points taking data at each point based on the test data selections. The details of how a test
runs are defined in the Test Definition (see Section 9). Certain aspects of how a test runs can be
customized using the
[Tools]
,
[Options]
,
[Run Test]
and
[End Test]
tabs (see Sections 11.6, 11.7).
End Test (see Section 7.2.3):
Presents the
<Test Notes>
screen to record test specific notes and
the
<Test Complete>
screen with various test conclusion options. Certain aspects of the test
conclusion can be customized using the
[Tools]
,
[Options]
,
[End Test]
tab (see Section 11.7).
7.2.1 TEST
INITIALIZATION
Test initialization is the first part of test execution. With the exception of Step below, all the
sequential steps of test initialization can be turned ON or OFF using the
[Tools]
,
[Options]
,
[Initialize Test]
tab (see Section 11.5). Test initialization proceeds as follows:
Initialize molbloc/molbox (see Section 7.2.1.1):
If the reference specified in the
System Configuration is a molbox,
COMPASS
loads the molbloc information into memory
and determines which features are supported by the molbox. Depending on the current setup,
this may take several seconds. This is the only Test Initialization step that cannot be disabled.
Select starting molbloc channel (see Section 7.2.1.2):
When a two channel molbox
reference is used, the initial molbloc channel can be selected if both channels have a
valid molbloc. If this step is turned OFF, the current molbloc channel is automatically used.
Select DUT (see Section 7.2.1.3):
The DUT selection tool is presented allowing
selection of the DUT Definition to run. If this step is turned OFF, the screen is not shown and
the last DUT used is automatically selected.
Summary of Contents for COMPASS FOR MOLBOX
Page 1: ...1998 2003 DH Instruments Inc COMPASS for molbox Flow Calibration Software User s Manual...
Page 16: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 2 N NO OT TE ES S...
Page 20: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 6 N NO OT TE ES S...
Page 64: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 50 N NO OT TE ES S...
Page 94: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 80 N NO OT TE ES S...
Page 112: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 98 N NO OT TE ES S...
Page 150: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 136 N NO OT TE ES S...
Page 166: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 152 N NO OT TE ES S...
Page 172: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 158 N NO OT TE ES S...
Page 192: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 178 N NO OT TE ES S...
Page 194: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 180 N NO OT TE ES S...
Page 200: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 186 N NO OT TE ES S...
Page 216: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 202 N NO OT TE ES S...
Page 220: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 206 N NO OT TE ES S...
Page 224: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 210 N NO OT TE ES S...
Page 226: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 212 N NO OT TE ES S...