COMPASS
for molbox™ User’s Manual
Page 81
©
1998-2003
DH Instruments, Inc
.
9
9
.
.
[
[
S
S
E
E
T
T
U
U
P
P
]
]
,
,
[
[
T
T
E
E
S
S
T
T
]
]
9.1 OVERVIEW
Test Definitions are used to define a specific test procedure including a sequence of flow points and a
large number of operational details. In
COMPASS [Run Test]
mode, a DUT to be tested is selected an
then the Test Definition to be executed is specified. A pointer to a default Test Definition for a DUT is
included in the DUT Definition
[Test]
tab (see Section 8.5.6).
Access the Test Definition Editor using the
[Setup], [Test]
menu option. When selected, the last
accessed Test Definition displays on the Test Definition Editor. While in the Editor, use the toolbar
options to create, edit, view and delete tests (see Section 9.5.1).
The Test Definition Editor is designed with the same editing concepts as the DUT Definition Editor
(see Section 8.5). Only one Test Definition can be edited at a time, however, other Definitions can be
viewed prior to completing the changes. The test scroll bar and the Test Selector are available to make
new test selections. A flashing indicator displays the ID number of the Test Definition currently being edited.
This ID will change as the test sort order changes (see Section 9.4). The edited test always displays in
blue text in the Test Definition Selector (see Section 9.4). Use these editing signals to locate the test that
is being edited if necessary.
Test Definitions can only be edited when COMPASS is not in an active run mode. If a run mode is active,
all Test Definition fields and selection functions are disabled.
9.2
CREATING TEST DEFINITIONS
Open the Test Definition Editor using the
[Setup], [Test]
menu option. Use the
<New>
test toolbar icon
to create a new Test Definition. Set up the test as desired then press the
<Save>
icon to store the test
information. Any conflicts that exist will cause an error message to display. Resolve the conflicts before
re-saving. The
<Restore>
icon is nonfunctional for new tests. The feature only restores to the last saved
state. Since new Test Definitions have not been saved, there is nothing to restore.
Familiarity with test execution sequence is necessary for proper creation of Test Definition
(review Section 7.1).
If a Test Definition exists that is similar to the one that will be created, use the <Copy> toolbar
feature to create a new Test Definition with the same characteristics. Then make the edits specific to
the new test. Don’t forget to change the <Record Label> prior to saving.
Summary of Contents for COMPASS FOR MOLBOX
Page 1: ...1998 2003 DH Instruments Inc COMPASS for molbox Flow Calibration Software User s Manual...
Page 16: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 2 N NO OT TE ES S...
Page 20: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 6 N NO OT TE ES S...
Page 64: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 50 N NO OT TE ES S...
Page 94: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 80 N NO OT TE ES S...
Page 112: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 98 N NO OT TE ES S...
Page 150: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 136 N NO OT TE ES S...
Page 166: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 152 N NO OT TE ES S...
Page 172: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 158 N NO OT TE ES S...
Page 192: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 178 N NO OT TE ES S...
Page 194: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 180 N NO OT TE ES S...
Page 200: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 186 N NO OT TE ES S...
Page 216: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 202 N NO OT TE ES S...
Page 220: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 206 N NO OT TE ES S...
Page 224: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 210 N NO OT TE ES S...
Page 226: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 212 N NO OT TE ES S...