COMPASS
for molbox™ User’s Manual
© 1998-2003
DH Instruments, Inc
.
Page 114
FEATURE
DESCRIPTION
Communications
Test
(action button)
This feature can be used only when the
<Output Type>
is RS-232 or IEEE-488.
Press this button to send the set
<Initialization Commands>
and a single target flow.
When the option is first selected, a dialog box prompts for the desired target flow.
A
<Spy Window>
will display with the command response information used
(see Section 6.7.7). This feature is designed to provide on the spot troubleshooting of
the remote command setup. If a command is not properly entered or the response
format is not correct, the problem should be visible in the spy window. Use the
<Remote Communications>
tool for further troubleshooting (see Section 13).
10.3.6 [FLOW
CONTROLLER]
TAB
The
[Setup]
,
[Hardware]
,
[Controller]
tab is used to manage the Hardware Definitions of
flow controllers that are included in the Test Definition
[Control]
tab,
<Controller>
drop down
list (see Section 10.2).
The
[Controller]
tab provides a list of the available flow controllers, displays complete information
on the selected controller, allows the controller to be edited or a new controller to be created.
To select a flow controller Hardware Definition to view, single click on the controller name and
serial number in the
<List>
display panel (upper left-hand corner). To edit an existing flow
controller, double click on the controller listing or click on the
<Edit>
toolbar function with the
desired controller highlighted. Double clicking the controller listing, clicking on the
<Edit>
toolbar function or clicking on the
<New>
toolbar function causes the
<Edit Flow Controller>
or
<Add Flow Controller>
pop-up to appear. These pop-ups are identical but differentiated
by name to make it clear which task is currently being performed.
The
<Edit Flow Controller>
or
<Add Flow Controller>
pop-ups allow a controller to be set
up from scratch or edited from an existing one.
The basic setup of the flow controller is very similar to setting up a DUT that is a
flow controller. The flow controller setup requires selections on three different tabs
<Range>
,
<Gas>
and
<Interface>
.
The purpose of the flow controller Definition fields and settings as well as instructions on how to
use them are provided in Table 39. When creating a new Definition, all the fields come up
blank. When editing an existing Definition, the fields come up populated with the information
on that device. When editing is complete, click
<OK>
to save changes.
Figure 65.
Hardware Definition,
[Controller]
Tab
Summary of Contents for COMPASS FOR MOLBOX
Page 1: ...1998 2003 DH Instruments Inc COMPASS for molbox Flow Calibration Software User s Manual...
Page 16: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 2 N NO OT TE ES S...
Page 20: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 6 N NO OT TE ES S...
Page 64: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 50 N NO OT TE ES S...
Page 94: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 80 N NO OT TE ES S...
Page 112: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 98 N NO OT TE ES S...
Page 150: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 136 N NO OT TE ES S...
Page 166: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 152 N NO OT TE ES S...
Page 172: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 158 N NO OT TE ES S...
Page 192: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 178 N NO OT TE ES S...
Page 194: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 180 N NO OT TE ES S...
Page 200: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 186 N NO OT TE ES S...
Page 216: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 202 N NO OT TE ES S...
Page 220: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 206 N NO OT TE ES S...
Page 224: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 210 N NO OT TE ES S...
Page 226: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 212 N NO OT TE ES S...