COMPASS
for molbox™ User’s Manual
Page 131
© 1998-2003
DH Instruments, Inc
.
Figure 74.
[Options], [End Test]
Tab!!!
Table 48. [Options]
,
[End Test]
Tab Fields
FEATURE
DESCRIPTION
Repeating final point of each
cycle
(radio button)
COMPASS
provides the capability to repeat test points using the
<Back>
<Run Toolbar>
button (see Section 6.5.1). However, if a test ends
automatically after the last point, there is no opportunity to repeat points when
the last point has completed. This options determines whether
COMPASS
will pause to allow points to be repeated after the last point is complete or
automatically end the test without pause.
There are three choices:
•
Enable repeat for fixed time delay:
The test will pause for a fixed time delay
after the last flow point is complete. During the delay, the
<Back>
<Run Toolbar>
button remains active allowing points to be repeated. This choice gives the
possibility of repeating points after the last point is complete but also causes the
test to continue automatically without operator intervention.
•
Prompt to repeat final point:
Systematically pauses the test at the end of each
test cycle and prompts the user as to whether the last test point should be
repeated. This selection guarantees that the operator will consider whether to
repeat points but it also prevents the test from completing automatically.
•
No final point verification:
The test will automatically conclude after the last
point. There is no opportunity to repeat points after the last point is complete. Use
this choice to prevent any delays or interruptions in the test sequence.
Prompt for test notes at the
end of the test
(check box)
If checked,
COMPASS
always prompts for test notes to be entered at the end
of the test.
Uncheck this option to avoid the test notes prompt. In this case, no test notes
can be included in the test’s Data File.
Auto generate report after
complete tests
(check box)
Check this option to automatically create a report at the end of complete
tests. The report is generated using the last report template used. The
<Test
Complete>
options still display at the end of the test.
Auto print generated report
(check box)
This option is enabled only when the
<Auto generate report after complete
test>
option is enabled. When this option is checked,
COMPASS
will
generate and print a report using the last selected report template.
Summary of Contents for COMPASS FOR MOLBOX
Page 1: ...1998 2003 DH Instruments Inc COMPASS for molbox Flow Calibration Software User s Manual...
Page 16: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 2 N NO OT TE ES S...
Page 20: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 6 N NO OT TE ES S...
Page 64: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 50 N NO OT TE ES S...
Page 94: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 80 N NO OT TE ES S...
Page 112: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 98 N NO OT TE ES S...
Page 150: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 136 N NO OT TE ES S...
Page 166: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 152 N NO OT TE ES S...
Page 172: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 158 N NO OT TE ES S...
Page 192: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 178 N NO OT TE ES S...
Page 194: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 180 N NO OT TE ES S...
Page 200: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 186 N NO OT TE ES S...
Page 216: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 202 N NO OT TE ES S...
Page 220: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 206 N NO OT TE ES S...
Page 224: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 210 N NO OT TE ES S...
Page 226: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 212 N NO OT TE ES S...