9.5.7 [LEAK
TEST]
TAB
The Test Definition
[Leak Test]
tab is used to specify whether a leak test will be run at the
beginning of the test (see Section 7.1) and/or after molbloc changes. The leak test used is the SYSTEM
leak test function
(see the molbox Operation and Maintenance manual, Leak Test Section).
The purpose of the
[Leak Test]
tab’s fields and settings as well as instructions on how to use
them are provided in Table 30.
Figure 53.
Test Definition Editor,
[Leak Test]
Tab
Table 30.
Test Definition Editor,
[Leak Test]
Tab Fields
FEATURE
DESCRIPTION
System leak
test
(check box)
Check this box to include a leak test at the beginning of the test. If a leak test is
run, results are included in the test’s Data File. Use the
<Abort test on failure>
to cause the test to abort after a leak test failure. Do not check this box if you
want the test to continue after a failed leak test.
Leak test after
molbloc
change
(check box )
Check this box for a leak test to be run after any molbloc change included in the test
point sequence. If the test fails, the failure is identified in the
<Data Grid><Status>
and in the test’s Data File (see Section 18.4). Use the
<Abort test on failure>
to determine whether the test should continue after a leak test failure.
Abort test on
failure
(check box)
Check this box to cause the test to abort if a leak test is failed. If not checked,
the leak check failure is recorded in the test’s Data File but the test continues. Note
that leaks in the flow system may cause the test data to be invalid.
9.5.8 [PURGE]
TAB
The Test Definition
[Purge]
tab is used to specify whether a gas purge routine will be run at the
beginning of the test (see Section 7.1) and/or after molbloc changes. The purge routine used is
the molbox purge function
(see the molbox Operation and Maintenance Manual, Purge Section).
The purpose of the
[Purge]
tab’s fields and settings as well as instructions on how to use
them are provided in Table 31.
Summary of Contents for COMPASS FOR MOLBOX
Page 1: ...1998 2003 DH Instruments Inc COMPASS for molbox Flow Calibration Software User s Manual...
Page 16: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 2 N NO OT TE ES S...
Page 20: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 6 N NO OT TE ES S...
Page 64: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 50 N NO OT TE ES S...
Page 94: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 80 N NO OT TE ES S...
Page 112: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 98 N NO OT TE ES S...
Page 150: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 136 N NO OT TE ES S...
Page 166: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 152 N NO OT TE ES S...
Page 172: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 158 N NO OT TE ES S...
Page 192: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 178 N NO OT TE ES S...
Page 194: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 180 N NO OT TE ES S...
Page 200: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 186 N NO OT TE ES S...
Page 216: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 202 N NO OT TE ES S...
Page 220: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 206 N NO OT TE ES S...
Page 224: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 210 N NO OT TE ES S...
Page 226: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 212 N NO OT TE ES S...