COMPASS
for molbox™ User’s Manual
© 1998-2003
DH Instruments, Inc
.
Page 14
5.1.1.1
SET UP AN RS-232 DUT
Use
[Setup]
,
[DUT]
to create a new DUT Definition (see Section 8). If a DUT
with similar characteristics already exists, it is easier to copy and edit the existing
DUT than to start from scratch. Before attempting to set up the RS-232 DUT,
refer to the DUT documentation to determine the commands used to measure
and control flow as well as the RS-232 interface settings. The first step is to enter
all of the DUT header information and decide whether to make this an
<Individual DUT>
or a
<DUT Profile>
. If the DUT is unique, use the
<Individual DUT>
option to
avoid the need to enter the DUT serial number and identification before starting a
test (see Section 8.5.2). Next, enter the flow range information. In most cases,
if a DUT outputs directly in RS-232 or IEEE-488, it will output in flow units.
However, this is not the case in this example. Note that
<Flow Units>
are
“sccm @ 0.00C” and
<Output Units>
are “%” (see Section 8.5.4).
Figure 1.
Example 1, DUT Range Setup
Select RS-232 as the DUT
<Output Type>
on the
[Read]
tab and enter the RS-232
settings for this DUT: 9600,E,7,1 (see Section 8.5.8). Then click the
<Edit>
button
next to the
<Read Commands>
list to display the remote command setup form.
To create the control code (02H), press the
<Insert Special Character>
button.
Select Hex then scroll to character “02” and press the insert button. The cursor
will turn into a pencil. Go to the first
<Remote Command>
entry field and click
the mouse in the left most position in the text display. A square or rectangle will
display where the mouse was clicked. Now enter the DUT flow command,
“01RFX” next to the control code. The DUT in this example outputs “N+24.55” as
24.55 %. So enter 1 in the
<Leading characters to strip>
. Press
<OK>
to save
the command setup and return to the DUT Definition Editor.
Summary of Contents for COMPASS FOR MOLBOX
Page 1: ...1998 2003 DH Instruments Inc COMPASS for molbox Flow Calibration Software User s Manual...
Page 16: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 2 N NO OT TE ES S...
Page 20: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 6 N NO OT TE ES S...
Page 64: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 50 N NO OT TE ES S...
Page 94: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 80 N NO OT TE ES S...
Page 112: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 98 N NO OT TE ES S...
Page 150: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 136 N NO OT TE ES S...
Page 166: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 152 N NO OT TE ES S...
Page 172: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 158 N NO OT TE ES S...
Page 192: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 178 N NO OT TE ES S...
Page 194: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 180 N NO OT TE ES S...
Page 200: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 186 N NO OT TE ES S...
Page 216: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 202 N NO OT TE ES S...
Page 220: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 206 N NO OT TE ES S...
Page 224: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 210 N NO OT TE ES S...
Page 226: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 212 N NO OT TE ES S...