COMPASS
for molbox™ User’s Manual
Page 15
© 1998-2003
DH Instruments, Inc
.
Figure 2.
Example 1, DUT Read Command Setup
Setting up the
<Set Point Type>
is similar to setting up the DUT
<Output Type>
.
On the
[Set]
tab, first select RS-232 as the
<Set Point Type>,
then enter the
RS-232 settings for the DUT: 9600,E,7,1. Then click the
<Edit>
button next to
the
<Set Commands>
list to display the remote command setup form. Insert the
(02H) control code as previously described. Now enter the DUT Set command,
“01STD” next to the control code. Instead of ## for the numerical set value,
either type or drag and drop the
<Set Unit Variable>
at the end of the set
command. This tells
COMPASS
where to place the numerical flow target value.
Save this DUT using the
<Save>
toolbar icon. The RS-232 communications port has
not been entered. This selection occurs as one of the last steps in test initialization,
<Setup DUT(s) - Data Acquisition>
(see Section 7.2.1.5).
5.1.1.2
SET UP A molbox AS THE REFERENCE
Use
[Setup]
,
[Hardware]
to open the Hardware Definition screen (see Section 10.3.3).
Select the
[Reference]
tab and check to see if the desired molbox is already in
the list. If so, verify that the remote settings are correct and close the display.
Otherwise press
<New>
to create a new reference (see Section 10.3.3). Select
DH Instruments
as the manufacturer and molbox1 (or RFM) as the model reference.
Then enter all other support information. For this example, the molbox will use
an IEEE-488 interface. The IEEE-488 address of the molbox can be determined
using the molbox’s front panel
[Special]
key (see the molbox Operation and
Maintenance Manual). In this example, 10 is used. Press
<OK>
to save this
Reference Definition.
Summary of Contents for COMPASS FOR MOLBOX
Page 1: ...1998 2003 DH Instruments Inc COMPASS for molbox Flow Calibration Software User s Manual...
Page 16: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 2 N NO OT TE ES S...
Page 20: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 6 N NO OT TE ES S...
Page 64: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 50 N NO OT TE ES S...
Page 94: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 80 N NO OT TE ES S...
Page 112: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 98 N NO OT TE ES S...
Page 150: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 136 N NO OT TE ES S...
Page 166: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 152 N NO OT TE ES S...
Page 172: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 158 N NO OT TE ES S...
Page 192: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 178 N NO OT TE ES S...
Page 194: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 180 N NO OT TE ES S...
Page 200: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 186 N NO OT TE ES S...
Page 216: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 202 N NO OT TE ES S...
Page 220: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 206 N NO OT TE ES S...
Page 224: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 210 N NO OT TE ES S...
Page 226: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 212 N NO OT TE ES S...