COMPASS
for molbox™ User’s Manual
© 1998-2003
DH Instruments, Inc
.
Page 116
Table 39.
Hardware Definition
, <Edit Controller>
Screen Fields
FEATURE
FIELD TYPE
DESCRIPTION
Manufacturer
Required Text Entry
Labels the device manufacturer. Write in a manufacturer name.
Model
Required Text Entry
Defines the device model. Write in a model name.
Serial Number
Required Text Entry
Defines the device serial number. Write in a serial number.
Identification
Optional Text Entry
Provides a device identification. Write in a device identification, if
desired. The field can be used for internal device tracking or any other
information desired. The value entered will be included in run test
Data Files and can be included in reports.
Do not use the following characters: \,/, :, *, ?, “, <,>, |, a
comma or tab character. These characters will cause problems
when importing the Data File or creating the Data File directory
based on the data directory naming convention selected using
the [Tools], [Options], [Data File] tab (see Section 11.8).
Calibration Date
Optional Text Entry
The value entered should represent the next required calibration
date for the device. No special formatting is expected or checked in
this field, however, entering the date in a consistent format including the
full year, month and day is recommended. The value entered is
logged in the run test Data Files and can be included in reports.
Set Range
Set Units
Required Drop
Down List Selection
Select the flow controller set point unit from the drop down list. If the
unit is not in the list, type the unit into the list box. If the current flow
units are selected in the list, the min/max set point fields are automatically
set to the min/max flow range fields. If the flow controller interface
is
<Voltage/Current>
, the set point unit must be V, mV, A or mA.
Set Range
Min Set
Required Numeric
Entry Field
Enter the lowest set point (set point corresponding to the
<Min
Flow>
) of the flow controller in the set point units specified
immediately above. If the set point units are the current flow units,
the
<Min Flow>
value is copied here and cannot be changed.
Set Range
Max Set
Required Numeric
Entry Field
Enter the highest set point (set point corresponding to the
<Max
Flow>
) of the flow controller in the set point units specified above.
If the set point units are flow units, the
<Max Flow>
value is copied
here and cannot be changed.
Flow Range Unit
Required
Drop
Down Selection List
Select the flow controller flow unit from the drop down list. The flow range
information entered should always be entered in the selected unit.
If the desired flow unit is not in the list, use the
<Edit Units>
option
at the top of the list to display the
<Flow Unit Editor>
(see Section 14).
All new units created automatically appear in the list of flow units.
Flow Range
Min Flow
Required Numeric
Entry
Enter the lowest flow controller flow. Typically this value is 0. In some special
cases, the minimum flow is non-zero. Regardless, the value entered
must be the expected flow when the flow controller’s
<Min Set>
occurs.
Flow Range
Max Flow
Required Numeric
Entry
Enter the maximum flow controller flow. This is the expected flow
when the flow controller’s
<Max Set>
occurs.
Calibration Gas
Optional
Drop
Down List
All molbox supported gases are selectable in this list. Select a gas
then enter a
<K Factor to N2>
(see immediately below) for all
gases supported by the flow controller. The
<K Factor to N2>
information is stored for every gas entered. 1 is used by default for
all gases not specifically set up.
Summary of Contents for COMPASS FOR MOLBOX
Page 1: ...1998 2003 DH Instruments Inc COMPASS for molbox Flow Calibration Software User s Manual...
Page 16: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 2 N NO OT TE ES S...
Page 20: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 6 N NO OT TE ES S...
Page 64: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 50 N NO OT TE ES S...
Page 94: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 80 N NO OT TE ES S...
Page 112: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 98 N NO OT TE ES S...
Page 150: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 136 N NO OT TE ES S...
Page 166: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 152 N NO OT TE ES S...
Page 172: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 158 N NO OT TE ES S...
Page 192: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 178 N NO OT TE ES S...
Page 194: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 180 N NO OT TE ES S...
Page 200: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 186 N NO OT TE ES S...
Page 216: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 202 N NO OT TE ES S...
Page 220: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 206 N NO OT TE ES S...
Page 224: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 210 N NO OT TE ES S...
Page 226: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 212 N NO OT TE ES S...