COMPASS
for molbox™ User’s Manual
Page 19
© 1998-2003
DH Instruments, Inc
.
5.1.3 EXAMPLE
#3
“How do I use my HP34401A DMM to measure the DUT output?”
The HP34401A is a popular high accuracy Hewlett Packard multimeter that has an IEEE-488 and
RS-232 interface. To read a voltage from this DMM, the command “MEAS:VOLT:DC?” must
be issued. This example demonstrates the basic requirements of setting up a DMM for DUT
data acquisition in the Hardware Definition
.
5.1.3.1
DUT DEFINITION SETUP
The only requirement for DUT to be tested using an external DMM is that the output type
of the DUT be
<Voltage/Current>
in the
[Read]
tab of the DUT Definition
(see Section 8.5.8).
5.1.3.2
DMM CONFIGURATION SETUP
Use
[Setup]
,
[Hardware]
, then select the
[DMM]
tab and click
<New>
to create
a new DMM (see Section 10.3.4). Enter the basic DMM setup information at the
top of the form. This particular DMM supports both RS-232 and IEEE-488 interfaces.
In this example IEEE-488 is used. If RS-232 were chosen, it is possible to use the
<Ref Com2>
option to use the COM2 port of the reference molbox to communicate
with the DMM. Select “Voltage (V)” from the
<Output Units>
field. Then press the
<Read Commands>
button. This displays the remote command Editor. Enter
“MEAS:VOLT:DC?” in the command field then press
<OK>
to store the setting
and return to the DMM Editor. Now is a good time to find the command to read
current so there is no need to re-configure this DMM in the future. Select “Current (A)”
from the
<Output Units>
field then again press the
<Read Command>
button.
Enter “MEAS:CURRENT:DC?” in the first field then press
<OK>
. This is the command
to read current in Amps should it be needed later. There are no leading characters
to strip and the command terminator is carriage line feed. Press
<OK>
to save this DMM. This Hardware Definition provides
COMPASS
with the
necessary information to use a DUT that outputs in V, mV and/or mA.
Figure 7.
Example 2, DMM Setup
Access the System Configuration using the
[Setup]
,
[System]
menu (see Section 10.2).
Select this DMM as the
<DUT Output DMM>
. Click
<Close>
to store the
configuration and return to the main program.
Summary of Contents for COMPASS FOR MOLBOX
Page 1: ...1998 2003 DH Instruments Inc COMPASS for molbox Flow Calibration Software User s Manual...
Page 16: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 2 N NO OT TE ES S...
Page 20: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 6 N NO OT TE ES S...
Page 64: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 50 N NO OT TE ES S...
Page 94: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 80 N NO OT TE ES S...
Page 112: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 98 N NO OT TE ES S...
Page 150: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 136 N NO OT TE ES S...
Page 166: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 152 N NO OT TE ES S...
Page 172: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 158 N NO OT TE ES S...
Page 192: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 178 N NO OT TE ES S...
Page 194: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 180 N NO OT TE ES S...
Page 200: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 186 N NO OT TE ES S...
Page 216: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 202 N NO OT TE ES S...
Page 220: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 206 N NO OT TE ES S...
Page 224: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 210 N NO OT TE ES S...
Page 226: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 212 N NO OT TE ES S...