COMPASS
for molbox™ User’s Manual
Page 41
© 1998-2003
DH Instruments, Inc
.
FEATURE
DESCRIPTION
Log 1 point at
fixed intervals
Logs an instantaneous data point at fixed time intervals. When the icon is pressed,
an input box displays to allow entry of the time interval. Then
COMPASS
will
automatically log a data point and repeatedly log data points until the
<Abort Data
Acquisition>
button is pressed or the run mode is closed. Taking data in this way
can be convenient when trying to monitor slowly evolving processes, i.e. the effect of
temperature on a DUT. Realize that taking data at short time intervals can result in a
very large data file.
The output of all manual entry devices must be entered for every data
point logged. This minimizes the automation provided by this function.
Average and log
points at fixed
intervals
Averages all device outputs and logs the results at fixed time intervals. When the
icon is pressed the
<Auto Point Average Setup>
(see Figure 20) displays to allow
entry of the averaging time and the intervals between averages. Then
COMPASS
will
automatically begin averaging data. When the average is complete, the results are
logged and the countdown to the next average begins. When the time specified as
the
<Restart average every (s)>
has been reached, another averaging cycle will
begin. This process is repeated until the
<Abort Data Acquisition>
button is
pressed or the run mode is closed. Taking data in this way can be convenient when
trying to mimic a test that includes multiple readings per point. Realize that taking
data at short time intervals can result in very large data file.
Press the [Abort Data Acquisition] option to abort fixed interval data
acquisition.
Figure 20.
<Auto Point Average Setup>
Abort data
acquisition
Press this icon to abort data acquisition at fixed intervals or a point currently being
averaged.
Never use other <Control Toolbar>
features when this icon is visible.
Changing the state of the system while a point is in progress leads to
faulty data logged in the data file.
Summary of Contents for COMPASS FOR MOLBOX
Page 1: ...1998 2003 DH Instruments Inc COMPASS for molbox Flow Calibration Software User s Manual...
Page 16: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 2 N NO OT TE ES S...
Page 20: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 6 N NO OT TE ES S...
Page 64: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 50 N NO OT TE ES S...
Page 94: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 80 N NO OT TE ES S...
Page 112: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 98 N NO OT TE ES S...
Page 150: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 136 N NO OT TE ES S...
Page 166: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 152 N NO OT TE ES S...
Page 172: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 158 N NO OT TE ES S...
Page 192: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 178 N NO OT TE ES S...
Page 194: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 180 N NO OT TE ES S...
Page 200: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 186 N NO OT TE ES S...
Page 216: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 202 N NO OT TE ES S...
Page 220: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 206 N NO OT TE ES S...
Page 224: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 210 N NO OT TE ES S...
Page 226: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 212 N NO OT TE ES S...