5.1.2 EXAMPLE
#2
Assuming the molbox1 from Example 1 is set up and it has the MFC control option,
“How do I calibrate a DUT that outputs 0 to 5 V corresponding to 0 to 1 slm?”
This example demonstrates the setup requirements of a common analog output DUT. The only
setup required to use the molbox MFC control option is to make sure the molbox is set up
with the proper remote interface in a Reference Hardware Definition and select the molbox in
the System Configuration (see Section 10).
5.1.2.1
SETUP <VOLTAGE/CURRENT> OUTPUT TYPE
Use the
[Setup]
,
[DUT]
menu choice to open the DUT Definition Editor (see Section 8).
Create a new DUT Definition and set up the basic header information corresponding
to the DUT. In the
[Range]
tab (see Section 8.5.4), select the flow range units
from the flow units list and enter the flow min and max values; 0 and 1 respectively.
Check the
<Controller>
checkbox on the
<Set Range>
panel. Now type or
select “V” as the output and set units. Enter 0 and 5 as the min and max output
and set point values. DUT’s that use voltage or current signals to indicate flow
should use the
<Voltage/Current>
<
Output Type>
. This tells
COMPASS
to use
the instrument currently selected as the
<DUT Output DMM>
in the System
Configuration to measure the DUT flow. Select
<Voltage/Current>
as the DUT
<
Set Mode>
. This tells
COMPASS
to use the
<Set Point Power Supply>
selected in the Hardware Definition to control the DUT flow. Use the
<Save>
toolbar icon to save the DUT.
5.1.2.2 REFERENCE
SETUP
Since we have assumed the reference molbox in Example #1 is already properly
set up, we only need to make the necessary Hardware Definition selections to
use the MFC control option. Use the
[Setup]
,
[System]
menu option to display
the
<System Config>
form (see Section 10.2). First select the molbox as the
<Reference>
. This automatically adds a
<Reference molbox>
option to the
<Set Point Power Supply>
,
<Set Point DMM>
and
<DUT Output DMM>
lists.
Select
<Reference molbox>
in all lists. If it were desired to set with the molbox but
measure with a DMM, the DMM should be selected in the
<DUT Output DMM>
list.
5.1.2.3
RUN THE TEST
Use
[Run]
,
[Run Test]
to start a test sequence. Select the DUT just created and
the desired Test Definition when prompted. There is no interface selection
required when setting up the DUT interface during the test initialization.
Summary of Contents for COMPASS FOR MOLBOX
Page 1: ...1998 2003 DH Instruments Inc COMPASS for molbox Flow Calibration Software User s Manual...
Page 16: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 2 N NO OT TE ES S...
Page 20: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 6 N NO OT TE ES S...
Page 64: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 50 N NO OT TE ES S...
Page 94: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 80 N NO OT TE ES S...
Page 112: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 98 N NO OT TE ES S...
Page 150: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 136 N NO OT TE ES S...
Page 166: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 152 N NO OT TE ES S...
Page 172: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 158 N NO OT TE ES S...
Page 192: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 178 N NO OT TE ES S...
Page 194: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 180 N NO OT TE ES S...
Page 200: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 186 N NO OT TE ES S...
Page 216: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 202 N NO OT TE ES S...
Page 220: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 206 N NO OT TE ES S...
Page 224: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 210 N NO OT TE ES S...
Page 226: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 212 N NO OT TE ES S...