Cradle
English
INTENDED USERS
Equipment should be used by clinicians medically trained in ultrasound imaging. User groups may include, but are
not limited to: Physicists, Radiation Oncologists, Surgeons and Urologists.
PERFORMANCE CHARACTERISTICS
n
Stepper allows for incremental longitudinal and rotational movement of the ultrasound transducer to achieve
its intended purpose.
n
Stepper includes a quick mounting, secure connection to CIVCO stabilizers to ensure stability of the
ultrasound transducer.
NOTE: For a summary of clinical benefits for this product, visit www.CIVCO.com.
CAUTION
Federal (United States) law restricts this device to sale by or on the order of a physician.
WARNING
n
Before use, you should be trained in ultrasonography. For instructions on the use of your transducer, see
your system's user guide.
n
Prior to use inspect device for signs of damage, if damage is evident do not use.
n
Cradle assembly is packaged non-sterile and is reusable. To avoid possible patient contamination, ensure
equipment is properly cleaned and disinfected before each use. See reprocessing section for instructions on
how to properly clean and disinfect.
n
If the product malfunctions during use or is no longer able to achieve its intended use, stop using the
product and contact CIVCO.
n
Report serious incidents related to the product to CIVCO and the competent authority in your Member State,
or appropriate regulatory authorities.
NOTE: Product is not made with natural rubber latex.
ATTACHING CRADLE
NOTE: For GE Healthcare ERB transducer: Transducer handle must be removed.
1. Open the cradle latch.
2. Place and orient transducer into the cradle.
3. Close cradle latch and tighten knob.
4. Loosen grid rail locking knobs and pull back on grid rail so grid platform is out of the way for initial
transducer positioning.
NOTE:
n
Free longitudinal movement achieved by fully disengaging longitudinal movement detent knob by
turning counterclockwise. Enable stepping function by turning knob clockwise until tight.
n
Free rotation of cradle achieved by disabling rotational movement detent knob by turning
counterclockwise. Enable rotational detent by turning knob clockwise until tight.
FUNCTIONAL PREUSE CHECKS
WARNING
n
Prior to each use perform the following checks to ensure optimal stepper performance.
1. Rotate cradle to ensure movement through full range of rotational travel.
2. If cradle does not move freely, apply lubricant that complies with hospital policies and procedures. Super
Lube
®
Multi-Purpose Synthetic Grease with Syncolon
®
(PTFE) is recommended.
3
Summary of Contents for CRADLE
Page 2: ......
Page 50: ...48 n n CIVCO www CIVCO com n n n n CIVCO n CIVCO GE Healthcare ERB 1 2 3 4 n n...
Page 75: ...n n CIVCO www CIVCO com n n n n CIVCO n CIVCO GE Healthcare ERB 1 2 3 4 n n 73...
Page 120: ...118 n n CIVCO www CIVCO com n n n n CIVCO n CIVCO GE Healthcare ERB 1 2 3 4 n n...
Page 163: ......