CTD-N Technical Manual
P/N 95F-6000-00 (February 2017)
page 147
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
C
Displays the current mode
Usage
C <CR> or <LF>
Operating Modes
Run, Open, Calibration
Description
The C command displays the current operating mode of the instrument.
Returns
If the instrument is in Open mode:
Open Mode
If the instrument is in Run mode:
Run Mode
If the instrument is in Calibration mode:
Cal Mode
See Also
***R, ***O, ***C
CH
Usage
CHx=ON <CR> or CHx=OFF<CR>
Operating Modes
Open
Description
For each command, typing the command by itself reports status, or the channel can be enabled (if
applicable) or disabled by typing CHx=ON or CHx=OFF. When the channel is disabled, its value is
excluded from the output string.
CH0 – Time
CH1 – Conductivity
CH2 – Temperature
CH3 - Pressure
CH4 - Salinity
CH5 - Sound Velocity