CTD-N Technical Manual
page 114
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
***E
Save the EE data base
Usage
***E <CR> or <LF>
Operating Modes
Open
Description
The ***E command causes all calibration constants and operational settings to be written to flash
memory.
CAUTION.
The ***E command overwrites the current calibration constants and operational
settings in flash memory. Before using the ***E command, use the RCAL command to verify
that the calibration constants are correct.
Returns
<CR><LF>
See Also
RCAL