CTD-N Technical Manual
page 110
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
***O
Go to the Open operating mode
Usage
***O <CR> or <LF>
Operating Modes
Run, Calibration
Description
The ***O command sets the instrument into the Open operating mode. In Open mode data are not
acquired and operational settings and calibration constants can be displayed and edited.
Returns
<CR><LF>
Entering a <CR> or <LF> character a second time and every time thereafter displays the following:
Open Mode