Step #3:
Thread the capillary assembly into the pres-
sure port slowly. You should see the oil wick
up into the capillary tube as you thread the
assembly into the end cap.
Have an absorbent paper towel on hand to
clean up any overflow that you may have.
Step #4:
Continue threading the assembly until the
threads are no longer exposed and the O-
ring is seated onto the end cap.
Have an absorbent paper towel on hand to
clean up any overflow that you may have.
Step #5:
Using hands only, tighten the assembly
onto the end cap to four inch-pounds.
***
HAND TIGHTEN ONLY!
***
*** MAXIMUM 4 INCH-POUNDS TORQUE
***
***
DO NOT OVERTIGHTEN
***
Have an absorbent paper towel on hand to
clean up any overflow that you may have.
If no oil is observed in the capillary tube re-
peat steps 1-5.
The installation is complete and the instru-
ment is ready to be deployed.